Abstract

Introduction Radiofrequency (RF)-induced tissue fusion shows great potential in sealing intestinal tissue without foreign materials. To improve the performance of RF-induced tissue fusion, a novel self-cooling jaw has been designed to minimize thermal damage during the fusion. Material and methods The prototype of self-cooling jaws was developed and manufactured. A total number of 60 mucosa-to-mucosa fusions were conducted using ex-vivo porcine intestinal segments with the proposed design and conventional bipolar jaws. The effects of intestinal fusion were evaluated based on temperature curves, burst pressure, thermal damage, and histological appearances. Results The self-cooling jaws showed significant decrease in temperature during the fusion process. An optimal burst pressure (5.7 ± 0.5 kPa) and thermal damage range (0.9 ± 0.1 mm) were observed when the applied RF power was 100 W. The thermal damage range of the prototype has almost decreased 36% in comparison with the conventional bipolar jaws (1.4 ± 0.1 mm). The histological observation revealed that a decrease of thermal damage was achieved through the application of self-cooling jaws. Conclusions The self-cooling jaws were proved to be effective for reducing the thermal damage during RF-induced tissue fusion, which could potentially promote the clinical application of tissue fusion techniques in the future.