Abstract

Secondary electrospray ionization high-resolution mass spectrometry (SESI-HRMS) is an emerging technique for the detection of volatile metabolites. However, sensitivity and reproducibility of SESI-HRMS have limited its applications in untargeted metabolomics profiling. Ion suppression in the SESI source has been considered to be the main cause. Here, we show that besides ion suppression, ion competition in the C-trap of Orbitrap instruments is another important factor that influences sensitivity and reproducibility of SESI-MS. Instead of acquiring the full mass-to-charge ratio (m/z) range, acquisition of consecutive m/z windows to minimize the ion competition effect allows the detection of more features. m/z window ranges are optimized to fill the C-trap either with an equal number of features or an equal cumulative intensity per window. Considering a balance between maximizing scanning speed and minimizing ion competition, splitting the m/z = 50–500 range into 4 windows is selected for measuring human breath and bacterial culture samples on SESI-Orbitrap MS, corresponding to a duty cycle of 2.3 s at a resolution of 140′000. In a small cohort of human subjects, the proposed splitting into 4 windows allows three times more features to be detected compared to the classical full m/z range method.