Abstract
The determination of trace and ultratrace elements in biological fluids, including urine and serum, by inductively coupled plasma mass spectrometry (ICP-MS) is discussed. Nonspectral interferences and their corrections by external calibration and calibrator addition are discussed in detail. External calibration with internal calibration and dilution is mostly sufficient to correct for encountered biological matrix effects. For some elements, such as Cs and Zn, the use of calibrator addition provides more accurate results. The importance of spectral interferences and their elimination by isotope selection was also studied. Two examples, Cu and Zn, demonstrate the prime importance of selecting an isotope with minimal polyatomic interferences for analysis. By using 65Cu and 68Zn, accurate results for urine and serum can be obtained without excessive pretreatment of samples. Two reference materials, Bio-Rad Lyphochek urine and Kaulson Contox sera, were analyzed. Accuracy was evaluated by comparison with target values, and precision was estimated by the CV within 95% confidence.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.