Abstract

Chemical interferences (ionization and oxide/hydroxide formation) on the atomic absorbance signal of lithium in FAAS analysis of brine samples are elaborated in this article. It is suggested that inadequate or overaddition of deionization buffers can lead to loss of sensitivities under particular operating conditions. In the analysis of brine samples, signal enhancing and oxide/hydroxide formation inducing signal reduction resulting from overaddition of deionization buffers can be seen with varying amounts of chemical buffers. Based on experimental results, the authors have arrived at the optimized operating conditions for the detection of lithium, under which both ionization and stable compound formation can be suppressed. This is a simplified and quick method with adequate accuracy and precision for the determination of lithium in routine brine samples from chemical plants or R&D laboratories, which contain comparable amounts of lithium with some other components.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.