Abstract

Ion suppression of drug response is a major source of imprecision for bioanalytical analysis using LC–MS/MS. Endogenous phospholipids cause ion suppression in both positive ESI and negative ESI modes and must be removed or resolved chromatographically. Three types of ion-exchange solid-phase extraction mediums were evaluated to determine their abilities to remove phospholipids. It was determined that although mixed mode phases fulfills the requirements of retaining both analytes and diverse metabolites, reverse phase retention mechanisms are detrimental in eliminating ion suppression caused by late eluting phospholipids. If an analyte and its metabolites can be retained using an ion-exchange mechanism alone, mixed mode extraction phases should be avoided.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.