Abstract
The minimization of copper and nickel interference on the determination of antimony by hydride generation atomic absorption spectrometry using picolinic acid and l-cysteine as masking agents was studied. Concentrations up to 4000 and 1000 mg l −1 of nickel and copper respectively were found to be tolerable in the presence of picolinic acid. The simultaneous presence of both transition metals could be tolerated by using a mixture of picolinic acid and l-cysteine. In the analysis of a nickel oxide certified reference material good agreement between the recommended and found values was verified. A detection limit (3 s, n=10) of 0.1 μg Sb g −1 in the original sample was estimated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.