Abstract
The authors describe a technique for minimally invasive anterior vertebroplasty for treating metastatic disease of the C-2 vertebra and discuss its application in 2 cases. After a 2-cm lateral neck incision is made, blunt dissection is performed toward the anterior inferior endplate of the C-2 vertebra. An 11-gauge needle is introduced through a tubular sheath and tapped into the inferior endplate of C-2, with biplanar fluoroscopy being performed to confirm position. The needle is subsequently advanced across the fracture line and into the odontoid process. Under fluoroscopic guidance, 2 ml of methylmethacrylate is injected into the odontoid process and vertebral body. This method is advantageous as 1) hyperextension of the neck is not performed, 2) the chance of inadvertent neurovascular or submandibular gland injury is minimized, 3) the possibility of cement leakage is decreased, and 4) hemostasis is better achieved under direct vision.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.