Abstract

To determine minimal endoillumination levels required to perform 3-dimensional heads-up vitreoretinal surgery and to correlate endoillumination levels used for measurements of heads-up display (HUD) luminous emittance. Prospective, observational surgical case series of 10 patients undergoing vitreoretinal surgery. Endoillumination levels were set to 40% of maximum output and were decreased at set intervals until the illumination level was 0%. Corresponding luminous emittance (lux) of the HUD was measured 40 cm from the display using a luxmeter (Dr. Meter, Model #LX1010BS). In 9 of 10 cases, the surgeon felt that they could operate comfortably at an endoillumination level of 10% of maximum output with corresponding HUD emittance of 14.3 ± 9.5 lux. In the remaining case, the surgeon felt comfortable at a 3% endoillumination level with corresponding HUD emittance of 15 lux. Below this threshold, subjective image dimness and digital noise limited visibility. Endoillumination levels were correlated with luminous emittance from the 3-dimensional HUD (P < 0.01). The average coefficient of variation of HUD luminance was 0.546. There were no intraoperative complications. With real-time digital processing and automated brightness control, 3-dimensional HUD platforms may allow for reduced intraoperative endoillumination levels and a theoretically reduced risk of retinal phototoxicity during vitreoretinal surgery.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.