Abstract
miniBELEN is a modular and transportable neutron moderated counter with a nearly flat neutron detection efficiency up to 10 MeV. Modularity implies that the moderator can be reassembled in different ways in order to obtain different types of response. The detector has been developed in the context of the Measurement of Alpha Neutron Yields (MANY) collaboration, which is a scientific effort aiming to carry out measurements of (𝛼, 𝑛) production yields, reaction cross-sections and neutron energy spectra. In this work we present and discuss several configurations of the miniBELEN detector. The experimental validation of the efficiency calculations using 252Cf sources and the measurement of the 27Al(𝛼, 𝑛) 30P reaction is also presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.