Abstract

The interest in implementation of miniaturized ultra-high performance liquid chromatography (UHPLC) in neurochemical research is growing because of the need for faster, more selective and more sensitive neurotransmitter analyses. The instrument performance of a tailor designed microbore UHPLC system coupled to electrochemical detection (ECD) is investigated, focusing on the quantitative monoamine determination in in vivo microdialysis samples. The use of a microbore column (1.0mm I.D.) requires miniaturization of the entire instrument, though a balance between extra-column band broadening and injection volume must be considered. This is accomplished through the user defined Performance Optimizing Injection Sequence, whereby 5μL sample is injected on the column with a measured extra-column variance of 4.5–9.0μL2 and only 7μL sample uptake. Different sub-2μm and superficially porous particle stationary phases are compared by means of the kinetic plot approach. Peak efficiencies of about 16000–35000 theoretical plates are obtained for the Acquity UPLC BEH C18 column within 13min analysis time. Furthermore, the coupling to ECD is shown suitable for microbore UHPLC analysis thanks to the miniaturized flow cell design, sufficiently fast data acquisition and mathematical data filtering. Ultimately, injection of in vivo samples demonstrates the applicability of the system for microdialysis analysis.