Abstract

A miniaturized detector consisting of three custom built hybrid circuits, a counter and a miniature high voltage power supply was designed to operate with custom synthesized Type Ib diamond crystals as sensors. Thick-film technology was incorporated in the circuit design. With a crystal having a volume of approximately 10 mm 3 and containing approximately 60 ppm paramagnetic nitrogen, the detector was capable of measuring γ-ray dose-rates as low as 7.5 μ Gy h −1. The response characteristic was linear up to 1 cGy h −1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.