Abstract

To address the increasing demand for sensitive and selective sample preparation methods for metal analysis; preconcentration of intended analyte from complex sample matrices before analysis is required to improve the performance of analysis instruments. In this study, we have engineered a sustainable and portable syringe-based hand-operable three-dimensionally (3D) printed sample pretreatment apparatus equipped with a replaceable bio-based thin- film metal sorbent. This device effectively addresses the challenges of sample matrix interference in metal analysis. A metal sorbent film composed of chitosan (CS) and polydopamine (PDA) leveraged the diverse functional groups in the CS/PDA matrix to significantly enhance the extraction efficiency for various metals. Our approach demonstrated excellent analytical performance, with coefficients of determination (R2) of 0.9982 for copper (Cu) and 0.996 for chromium (Cr). The method achieved low limits of detection (LOD) of 0.3 μg L−1 for Cr and 0.7 μg L−1 for Cu. Precision and practicality assessments using actual urine samples yielded satisfactory relative standard deviations (RSD%) ranging from of 1.6 %–8.5 % for both metals, indicating minimal interference from the sample matrix. Moreover, our approach exhibited robust performance even after seven consecutive extraction and desorption cycles, highlighting its sustainability and practical applicability for laboratory and on-site sample pretreatment.