Abstract

Abstract The construction and optimization of five new types of miniaturized flow-through electrolytic cells with lead cathode and platinum anode for electrochemical hydride generation in atomic absorption spectrometry (HG-QFAAS) were achieved during this research study. The ion-exchange membrane was not part of these cells and only one carrying electrolyte for both electrode chambers was used. Hydride generation efficiency achieved was either comparable or higher than the one recorded for the classic thin-layer generation cell. The inner volume of the cathode chamber was reduced to a quarter of the classic thin-layer flow-through cell. Compared to the commonly used thin-layer flow-through cell, higher sensitivity (7.32×103 dm3 µg−1) and better limit of detection (0.32 µg dm−3) were obtained for selenium determination using two of these new generators.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.