Abstract

This paper presents a new strategy of packaging developed for medical applications which includes a sensor and an integrated circuit inside a hermetic silicon box that could be embedded in a cardiac lead in order to monitor the endocardial acceleration. The electronic components are placed on a silicon interposer wafer which was bonded by the eutectic AuSi with a silicon wafer lid containing cavities. Different metal stacks of the sealing ring including a barrier and Au have been studied. The gas content and hermeticity of the package were analyzed using Residual Gas Analysis (RGA) and biodegradation were tested in saline solution. The final silicon package was encapsulated with biocompatible materials that have high conformality deposition and act as good bi-directional barrier. Materials behavior that had been reported in literature as biocompatible and compatible with fabrication in standard clean rooms were studied. Finally, packaging materials were tested in cytotoxicity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.