Abstract

An innovative design for a miniature time-of-flight mass spectrometer has been developed employing several newly designed components. These include: (1) a gridless, focusing ion source allowing for the use of very high extraction energies in a maintenance-free design, (2) a new method of construction for an ion reflector using rolled flexible circuitboard material, and (3) an improved center-hole microchannel plate detector assembly that significantly reduces the noise (or 'ringing') inherent in the coaxial design. A prototype time-of-flight instrument was constructed and used to evaluate the performance of these components. Compared to previous designs, results indicate that background noise for data acquired on this instrument is substantially reduced, resolution is improved, and the potential for mass producing this instrument in an inexpensive and rugged package for field-portable and remote installations is significantly enhanced. Copyright 2000 John Wiley & Sons, Ltd.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.