Abstract

Recent interest in small-scale flow devices has created the need for miniature instruments capable of measuring scalar flow properties with high spatial resolution. We present a miniature rainbow schlieren deflectometry system to nonintrusively obtain quantitative species concentration and temperature data across the whole field. The optical layout of the miniature system is similar to that of a macroscale system, although the field of view is smaller by an order of magnitude. Employing achromatic lenses and a CCD array together with a camera lens and extension tubes, we achieved spatial resolution down to 4 mum. Quantitative measurements required a careful evaluation of the optical components. The capability of the system is demonstrated by obtaining concentration measurements in a helium microjet (diameter, d=650 microm) and temperature and concentration measurements in a hydrogen jet diffusion flame from a microinjector (d=50 microm). Further, the flow field of underexpanded nitrogen jets is visualized to reveal details of the shock structures existing downstream of the jet exit.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call