Abstract
A plane-parallel-plate electron energy analyzer is described based on the second-order-focusing 30° injection geometry. The small size of the analyzer permits it to be attached to a commercial rotary feedthrough and to be used in angle-resolved photoemission studies. The analyzer is operated in a constant pass energy mode, and the electron energy spectrum is obtained by sweeping the voltage on a preretardation/acceleration stage. A resolution function ΔE/E equal to ∼6% leads to a resolution of lower than 0.2 eV at sufficiently low pass energies. The performance is illustrated with results obtained using synchrotron radiation in a system having two identical analyzers.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.