In the Micra Transcatheter Pacing Study, a novel leadless pacemaker system delivered using a transcatheter system, was tested for strength duration and battery longevity. Despite its size, the electrical performance of the device was comparable with current ventricular demand pacemakers and utilization of 0.24 vs 0.4 milliseconds pulse width increased the projected battery longevity by 0.83 years.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.