Abstract

We have investigated the application of ultrasound modulated coherent light for quantitative determination of the ratio of dye concentrations and total concentration of absorbers in a blood vessel-mimicking sample. A 3-mm-diam tube containing the mixture of dyes inside an Intralipid-based gel with optical properties similar to tissue was interrogated by two different laser wavelengths in combination with intense microsecond ultrasound bursts. The use of calibration curves allowed us to extract quantitative information on the ratio of dye concentrations with the accuracy of better than 15%, as well as on the total concentration. Furthermore, we demonstrated the feasibility to obtain a quantitative 3-D map of the absorbing structure with a spatial resolution of better than 3 mm. These findings give an outlook to apply this technique for noninvasive 3-D mapping of oxygen saturation and total concentration of hemoglobin in tissue.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.