We report a retrospective study on 122 THA using Duraloc "1200 series" cups. All patients received a 28-mm head (107 Al2O3, 15 CoCr). The mean age of the patients at surgery was 56.3 years and the mean follow-up was 137 months.Mean migration at two years using the EBRA method was 0.18 mm cranially and 0.16 mm medially. Mean polyethylene wear was 0.03 mm/year using the Imagika software and 0.06 mm/year using EBRA. Only one patient presented with abnormal polyethylene wear of 1.7 mm at 90 months with Imagika and 2.4 mm with EBRA. The Duraloc cup seems to be a reliable implant in respect of bone fixation and resistance to wear.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.