Abstract

For anthropomorphic phantom measurements differences between calculated and measured entrance-exit doses of head, neck, shoulder, lung and thick pelvis were 0.8%, 2.7%, 26.4%, 4.4% and 4.9% and for midline doses were 1.6%, 1.6%, 6.3%, -1.4% and 7.4% respectively. For patients; TLD differences were within -4.13% ile 6.7%, -3.3% ile 3.9%, 5.1% ile 16.6%, -7.8% ile 2.4%, and 3.6% ile 7.1% respectively. For thick pelvis measurements with ion chamber differences were within %0.1-1.9.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.