Abstract
Free-running thulium laser pulses (Cr:Tm:YAG, lambda = 2.01 microm, tp = 300 micros) were applied to a purified, degassed water sample and the resulting temperature rise was investigated by an optical temperature probe. The probe detected water reflectance index changes with temperature and also the onset of vaporization, which was found to occur in a superheat regime, at approximately 230 degrees C. The experimental data were compared with theoretical temperature calculations, and deviations of less than 20 degrees C were stated. The best agreement between theory and experiment was found for temperatures below 180 degrees C, defining by this the method's high accuracy limit. In conclusion, both the optical temperature probe and the presented calculations can help to improve dosimetry in pulsed IR laser applications by precise temperature measurement and prediction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.