Abstract
To determine how the ear-canal sound pressures generated by earphones differ between normal and pathologic middle ears. Measurements of ear-canal sound pressures generated by the Etymtic Research ER-3A insert earphone in normal ears (N = 12) were compared with the pressures generated in abnormal ears with mastoidectomy bowls (N = 15), tympanostomy tubes (N = 5), and tympanic-membrane perforations (N = 5). Similar measurements were made with the Telephonics TDH-49 supra-aural earphone in normal ears (N = 10) and abnormal ears with mastoidectomy bowls (N = 10), tympanostomy tubes (N = 4), and tympanic-membrane perforations (N = 5). With the insert earphone, the sound pressures generated in the mastoid-bowl ears were all smaller than the pressures generated in normal ears; from 250 to 1000 Hz the difference in pressure level was nearly frequency independent and ranged from -3 to -15 dB; from 1000 to 4000 Hz the reduction in level increased with frequency and ranged from -5 dB to -35 dB. In the ears with tympanostomy tubes and perforations the sound pressures were always smaller than in normal ears at frequencies below 1000 Hz; the largest differences occurred below 500 Hz and ranged from -5 to -25 dB. With the supra-aural earphone, the sound pressures in ears with the three pathologic conditions were more variable than those with the insert earphone. Generally, sound pressures in the ears with mastoid bowls were lower than those in normal ears for frequencies below about 500 Hz; above about 500 Hz the pressures showed sharp minima and maxima that were not seen in the normal ears. The ears with tympanostomy tubes and tympanic-membrane perforations also showed reduced ear-canal pressures at the lower frequencies, but at higher frequencies these ear-canal pressures were generally similar to the pressures measured in the normal ears. When the middle ear is not normal, ear-canal sound pressures can differ by up to 35 dB from the normal-ear value. Because the pressure level generally is decreased in the pathologic conditions that were studied, the measured hearing loss would exaggerate substantially the actual loss in ear sensitivity. The variations depend on the earphone, the middle ear pathology, and frequency. Uncontrolled variations in ear-canal pressure, whether caused by a poor earphone-to-ear connection or by abnormal middle ear impedance, could be corrected with audiometers that measure sound pressures during hearing tests.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.