Abstract

Ventricular backup leads may be considered in selected patients with His bundle pacing (HBP), but it remains unknown to what extent this is useful. A total of 184 HBP patients were studied. At last follow-up, 147 (79.9%) patients retained His bundle capture at programmed output. His bundle pacing lead revision was performed in 5/36 (13.9%) patients without a backup lead and in 3/148 (2.0%) patients with a backup lead (P = 0.008). One patient without a backup lead had syncope due to atrial oversensing. Thus, implantation of ventricular backup leads may avoid lead revision and adverse events in selected HBP patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.