Abstract

This paper explores the use of a microwave-reflectometry-based system for the automatic control and real-time monitoring of the flow and of the liquid level in intravenous (IV) medical infusions. In medical and hospital contexts, other kinds of devices, mainly based on the optical detection and counting of the infusion drops, are used. Nevertheless, the proposed system is aimed at circumventing some typical drawbacks deriving from the adoption of these traditional methods, thus allowing an efficient alternative for automatically monitoring the instantaneous flow of IV medical solutions. To this purpose, the proposed system combines microwave time-domain reflectometry (TDR) measurements with a noninvasive sensing element (i.e., strip electrodes directly attached to the external surface of the infusion bottle). Experimental results confirm that, by using low-cost portable TDR devices, the solution flow process can be controlled with acceptable accuracy. Therefore, the proposed method can be regarded as a promising control tool for in-hospital patient management as well as for telemedicine programs.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call