Abstract

An antibody-detecting sensor is described that is based on a microwave electrodynamic resonator. A polystyrene film with immobilized bacteria deposited on a lithium niobate plate was placed at one end of the resonator and was used as the sensing element. The second end was electrically shorted. The frequency and depth of the reflection coefficient S11 for three resonances in the range 6.5–8.5 GHz were used as an analytical signal to examine antibody interactions with bacteria and determine the time required for cell immobilization. The sensor distinguished between situations in which bacteria interacted with specific antibodies and those in which no such interaction occurred (control). Although the cell–antibody interaction changed the frequency and depth of the second and third resonance peaks, the parameters of the first resonance peak did not change. The interaction of cells with nonspecific antibodies did not change the parameters of any of the peaks. These results are promising for use in the design of methods to detect specific antibodies, which can supplement the existing methods of antibody analysis.