Abstract

This paper explores the competency of the time domain ultra-wideband (UWB)-circular synthetic aperture radar (CSAR) to image the breast and detect tumors. The image reconstruction is performed using a time domain global back projection technique adapted to the circular trajectory data acquisition. This paper also proposes a sectional image reconstruction method to compensate for the group velocity changes in different layers of a multilayer medium. Experiments on an advanced breast phantom examines the suitability of this technique for breast tumor imaging. The advanced breast phantom is designed based on a MRI of a real patient, fabricated using 3D printing technology, and filled with liquids that emulate normal and cancerous tissues. The measurement results, compared with MRI imaging of the phantom, demonstrate the suitability of the UWB-CSAR method for breast tumor imaging. This method can be a tool for early diagnosis as well as for treatment monitoring during chemotherapy or radiotherapy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.