Abstract

The analysis of trace elements in biological samples is essential to extend our knowledge on human health and disease. Inductively coupled plasma mass spectrometry (ICP-MS) makes it possible to simultaneously determine these elements in trace amounts. Before analysis, however, biological samples such as organs and tissues must be liquefied and extra organic materials must be decomposed by acid digestion. We established a method of microwave digestion using dual PTFE containers to minimize the amount of samples. Samples (35-45 mg) of standard reference materials, bovine liver (1577a, NIST) and fish flesh (MA-A-2, IAEA), were weighed in PTFE-PFA vials and a small amount of nitric acid (0.5 ml) was added. The vials were sealed and two PTFE-PFA vials were placed in a PTFE-TFM vessel containing 6 ml of pure water. Then the vessels were placed in a rotor and the samples were digested for 38 min in a microwave oven according to a pre-set program. After the program was completed, the samples were analyzed by ICP-MS. The determined values of elements of the microwave-digested samples matched the certified values of the standard reference materials. Therefore, the digestion using dual containers was successfully applied to small samples.