Abstract
According to Valderrama (1981) aquatic samples are successfully digested with peroxodisulphate prior to total nitrogen determination. This study compares the utilization of a microwave oven and an autoclave for the preparation of liquid samples in terms of precision and accuracy of the obtained recovery. The comparison encompasses both reference solutions such as urea, amino acids, nucleic acids and recalcitrant compounds as well as natural water samples. Although the observed differences are small, the microwave digestion technique proofs significantly more precise. On the other hand, biases introduced by each of the techniques are similar. The slightly but significantly better precision and its easy utilization on board research vessels make the microwave digestion technique highly recommendable.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Water Research
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.