Abstract

The use of microwave-assisted solvent extraction (MASE) as an alternative for conventional solvent extraction procedures for the determination of some triazine herbicides in soil samples has been investigated. In this study MASE method development was focused on the selection of a suitable extraction solvent prior to the instrumental analysis of uncleaned extracts with gas chromatography and nitrogen-specific detection. A mixture of dichloromethane-methanol (90∶10, v/v) yielded recoveries ranging from 89 to 103 (spiked level 200 μg/kg) with RSDs ranging from 2.1 to 5.3%. This solvent mixture is also very convenient for further procedure. The selected MASE procedure was tested by analyzing freshly spiked soil samples and samples with aged residues of atrazine, desethylatrazine, desisopropylatrazine and simazine. The results were compared with those obtained by a conventional liquid extraction method. The comparative study indicated that MASE yields recoveries at least as good as those obtained by the conventional method. Moreover, the MASE procedure provides low solvent consumption in combination with a high sample throughput.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.