Abstract

This study investigated a microwave-assisted aqua regia digestion of catalyst samples prior to the determination of Pt, Pd, Rh and Pb by ICP-OES. While optimising the method, the effect of digestion temperature on the digestion efficiency was investigated. Few aqua regia digestions were also carried out using a high pressure asher, with a digestion temperature of 280°C. In addition, XRD (X-ray diffractometry) was used to characterise digestion residue. The study showed that accurate and reliable results for Pt, Pd, Rh and Pb in a recycled monolith catalyst (NIST SRM 2557) were obtained by ICP-OES after the microwave assisted aqua regia digestion. The amounts of Pt, Pd and Rh in the prepared catalyst samples were also at the expected levels. Furthermore, the results indicated that the main part of the impurities and supporting material elements can be quantitatively extracted from the catalyst samples by using the microwave-assisted aqua regia sample pre-treatment.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call