Abstract

OBJECTIVE. The objective of our study was to evaluate the clinical performance of a new high-frequency (HF) microwave ablation (MWA) technology with spatial energy control for treatment of lung malignancies in comparison with a conventional low-frequency (LF) MWA technology. MATERIALS AND METHODS. In this retrospective study, 59 consecutive patients (mean age, 58.9 ± 12.6 [SD] years) were treated in 71 sessions using HF spatial-energy-control MWA. Parameters collected were technical success and efficacy, tumor diameter, tumor and ablation volumes, ablation time, output energy, complication rate, 90-day mortality, local tumor progression (LTP), ablative margin size, and ablation zone sphericity. Results were compared with the same parameters retrospectively collected from the last 71 conventional LF-MWA sessions. This group consisted of 56 patients (mean age, 60.3 ± 10.8 years). Statistical comparisons were performed using the Wilcoxon-Mann-Whitney test. RESULTS. Technical success was 98.6% for both technologies; technical efficacy was 97.2% for HF spatial-energy-control MWA and 95.8% for LF-MWA. The 90-day mortality rate was 5.1% (3/59) in the HF spatial-energy-control MWA group and 5.4% (3/56) in the LF-MWA group; for both groups, there were zero intraprocedural deaths. The median ablation time was 8.0 minutes for HF spatial-energy-control MWA and 10.0 minutes for LF-MWA (p < 0.0001). Complications were recorded in 21.1% (15/71) of HF spatial-energy-control MWA sessions and in 31.0% (22/71) of LF-MWA sessions (p = 0.182); of these complications, 4.2% (3/71) were major complications in the HF spatial-energy-control MWA group, and 9.9% (7/71) were major complications in the LF-MWA group. The median deviation from ideal sphericity (1.0) was 0.195 in the HF spatial-energy-control MWA group versus 0.376 in the LF-MWA group (p < 0.0001). Absolute minimal ablative margins per ablation were 7.5 ± 3.6 mm (mean ± SD) in the HF spatial-energy-control MWA group versus 4.2 ± 3.0 mm in the LF-MWA group (p < 0.0001). In the HF spatial-energy-control MWA group, LTP at 12 months was 6.5% (4/62). LTP at 12 months in the LF-MWA group was 12.5% (7/56). Differences in LTP rate (p = 0.137) and time point (p = 0.833) were not significant. CONCLUSION. HF spatial-energy-control MWA technology and conventional LFMWA technology are safe and effective for the treatment of lung malignancies independent of the MWA system used. However, HF spatial-energy-control MWA as an HF and high-energy MWA technique achieves ablation zones that are closer to an ideal sphere and achieves larger ablative margins than LF-MWA (p < 0.0001).

Highlights

  • Complications were recorded in 21.1% (15/71) of HF spatial-energy-control microwave ablation (MWA) sessions and in 31.0% (22/71) of LF-MWA sessions (p = 0.182); of these complications, 4.2% (3/71) were major complications in the HF spatial-energy-control MWA group, and 9.9% (7/71) were major complications in the LF-MWA group

  • HF spatial-energy-control MWA as an HF and high-energy MWA technique achieves ablation zones that are closer to an ideal sphere and achieves larger ablative margins than LF-MWA (p < 0.0001)

  • Surgery is the treatment of choice for curative therapy of primary lung malignancies, but most patients are ineligible for surgery because of poor cardiopulmonary reserve, unfavorable tumor localization, too many nodules, or systemic metastatic disease [3,4,5]

Read more

Summary

Methods

59 consecutive patients (mean age, 58.9 ± 12.6 [SD] years) were treated in 71 sessions using HF spatial-energy-control MWA. Results were compared with the same parameters retrospectively collected from the last 71 conventional LF-MWA sessions This group consisted of 56 patients (mean age, 60.3 ± 10.8 years). Technical success was defined as correct accomplishment of ablation according to protocol and coverage of the complete index tumor with the ablation zone as shown on CT performed immediately at the end of intervention. Technical efficacy was defined as complete coverage of the index tumor with ablation zone on contrast-enhanced CT 24 hours after ablation [16, 28]. LTP was defined as a tumor focus connected to the edge of an ablation zone that was previously considered technically efficient on CT performed 24 hours after the ablation. Recurrences were evaluated on contrast-enhanced CT 3 months after ablation

Results
Discussion
Conclusion