Abstract

In this paper, biomedical porous Ti–Nb–Ta–Zr with 40% porosity and 166 ± 21 μm macro-pore size was successfully fabricated by space holder method. The microstructure, Vickers hardness, compressive and electrochemistry behavior were studied. It results that a few second phases exist in β matrix of the porous Ti–Nb–Ta–Zr. Its Young's modulus is 0.8 GPa, close to 0.01–3 GPa for trabecular bone. The total recovery strain ratio and pseudoelastic strain ratio are 8.8% and 2.7%, respectively. It fails mainly by brittle cleavage with the fan-shaped and smooth cleaved facets. Although, local ductile fracture by a few dimples and a small amount of transcrystalline fracture with the cleavage of similarly oriented laths in a colony are observed on the fracture surface. The impedance spectrum of porous Ti–Nb–Ta–Zr has the characteristics of half capacitive arc resistance, showing good corrosion resistance in SBF solution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.