Abstract
The effect of fast electron beam intensity fluctuations and of the dead time of the electron detectors on the shape of TDC spectra in tagged photon experiments is explained. Some consequences for the data analysis and for the use of multi-hit TDCs are also discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.