Abstract
Total knee replacement implants consisting of a Co–29Cr–6Mo alloy femoral component and a Ti–6Al–4V tibial component are the basis for the additive manufacturing of novel solid, mesh, and foam monoliths using electron beam melting (EBM). Ti–6Al–4V solid prototype microstructures were primarily α -phase acicular platelets while the mesh and foam structures were characterized by α ′ -martensite with some residual α . The Co–29Cr–6Mo containing 0.22% C formed columnar (directional) Cr 23C 6 carbides spaced ∼2 μm in the build direction, while HIP-annealed Co–Cr alloy exhibited an intrinsic stacking fault microstructure. A log–log plot of relative stiffness versus relative density for Ti–6Al–4V and Co–29Cr–6Mo open-cellular mesh and foams resulted in a fitted line with a nearly ideal slope, n = 2.1 . A stress shielding design graph constructed from these data permitted mesh and foam implant prototypes to be fabricated for compatible bone stiffness.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the Mechanical Behavior of Biomedical Materials
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.