Synopsis The authors present a histologic analysis of 19 Branemark titanium implants retrieved for different causes: four implants were removed for abutment fracture, one for dental nerve dysesthesia, two for bone overheating, two for peri-implantitis, nine for mobility, one for unknown causes. In the implants removed for fracture a high bone-implant contact percentage was present (71.83±4.96%) with compact, mature bone at the interface. The picture of the failure due to bone overheating was characteristic with the presence of bone sequestra and of a gap between implant and bone filled by lymphocytes and plasma cells: many bacteria surrounded the necrotic bone and no newly regenerated bone was present. In peri-implantitis an inflammatory infiltrate was observed in the peri-implant tissues: a dense fibrous connective tissue was present around implants failed for mobility. The microscopical picture is certainly extremely important in identifying the causal determinants of an implant failure.