Abstract
Programmatic operations at the Los Alamos National Laboratory Plutonium Facility (TA-55) involve working with plutonium and other highly toxic, alpha-emitting materials. The spread of radiological contamination on surfaces, airborne contamination, and excursions of contaminants into the operator's breathing zone are prevented using gloveboxes. The TA-55 Glovebox Safety Program augments this passive safety feature. A key element of this program is to perform microscopic analyses of glovebox glove failures and breaches when their causes are unknown. Initial examinations provide little information about the causes of failures and breaches. This has prompted the development of an instrument that permits cross-sections of glove perforations to be studied. This paper discusses the methodology used to perform microscopic analysis of failures and breaches. It presents three case studies in which microscopic analyses have helped to determine the causes of the glove openings.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.