Abstract
A simple, semi-automated, microcolumn solid-phase extraction (SPE) system is optimized for the extraction, preconcentration and HPLC analysis of seven different explosives and explosive derivatives contaminating seawater, river water and well water samples. The microcolumns were constructed from 1/16 in. O.D. PTFE tubing (1 in.=2.54 cm) packed with 0.5–1.5 mg of SPE material, LiChrolut EN or Porapak R. The extraction system consisted of two syringe pumps and several solenoid valves. Optimal detection limits were realized when the sample water flow-rate was maximally increased within the limits of the pump, 5–10 ml/min (despite exceeding the breakthrough threshold of the SPE microcolumn), and when the eluate volume collected from the column was minimized, <5 μl (despite very low recovery percentages).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.