Abstract
We developed a miniaturized LC-MS system with a high-recovery phosphopeptide enrichment protocol that allows phosphoproteome analysis of 10(4) cells. In the enrichment protocol, the key step is to add sodium deoxycholate and sodium lauroyl sarcosinate to the buffer solution for protein extraction and digestion and to omit any subsequent desalt/desurfactant step before phosphopeptide enrichment. The phosphopeptides enriched by hydroxy acid-modified metal oxide chromatography (HAMMOC) are directly injected onto a miniaturized LC column using a nitrogen-pressure-driven cell, instead of switching valve-type injectors. The miniaturized analytical column of 25 μm diameter provided a 3.6-fold improvement in sensitivity over the conventional 100 μm diameter column. Overall, our analytical system provided approximately 80-fold improvement on average in the LC-MS response, and we identified 1011 unique phosphorylated sites based on 995 unique phosphopeptides from a single analysis of 10(4) HeLa cells (approximately 1 μg of proteins). This is the most sensitive phosphoproteomics system that has so far been reported for proteome-wide analysis of in vivo phosphorylation in mammalian cells.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.