Abstract

The methods of uniform heating and resistive (Joule) heating for microscale freestanding surface-micromachined thin metal film specimens were evaluated by a combination of full-field strain measurements by optical microscopy/Digital Image Correlation (DIC) and microscopic infrared (IR) imaging. The efficacy of each method was qualitatively and quantitatively evaluated with the aid of strain fields and IR-obtained temperature distributions along 850 nm thick freestanding microscale specimens subjected to uniaxial tension while heated by each method. The strain and temperature fields were quite uniform in experiments carried out with uniform specimen heating except for minor end-effects at the specimen grips. However, the resistively heated specimens showed highly uneven temperature distribution varying by 50°C along the 1,000 μm specimen gauge length. This high temperature gradient resulted in strain localization and 40% reduction in yield and ultimate tensile strengths of resistively heated specimens compared to the uniformly heated ones. Therefore, it is concluded that resistive heating is not a reliable method for conducting microscale temperature experiments with metallic films.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.