Abstract

Microtiter plate methods were developed for the enzymatic determination of serum total cholesterol (TC), high density lipoprotein cholesterol (HDL-C) and triglyceride (TG), and for the turbidometric determination of apolipoproteins. The micromethods resulted in accurate, precise values that were in good agreement with the conventional spectrophotometric assays. The coefficient of variation for TC determinations was 4.5% or less and bias was 5% or less. The lipid micromethod assays are sensitive to 10 mg/dL or less, and the apolipoprotein assay to 1 mg/dL. Less than 100 microL of serum suffices for TC, TG and apoprotein assays; HDL-C requires an additional 100 microL of serum. Advantages of the micromethods include reductions in assay time and in the amount of reagents required.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.