Abstract
Micropipette tip (tip) pipetting accuracy and precision are functions of tip manufacture consistency, morphological variation, and retained inner tip wall particulate matter. Irregular tip inner wall surfaces and retained particulate matter cause pipetting inaccuracy. Washed and unwashed tips from seven manufacturers were compared using weight by difference (six), microscopic Luxol Fast Blue staining (seven), and matrix-assisted laser desorption/onization time-of-flight (MALDI-TOF) mass spectrometry (six). Photomicrographs revealed tip irregularity and inner wall retained particulate matter. Computer model analysis was used to identify tip irregularity and retained matter. These tests established the utility of a pipette solvent wash to increase the performance and the accuracy of tips and, thus, improve the MALDI mass spectra obtained.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.