Abstract

A family of U-tube manometers, suitable for use as standard instruments, has been designed and constructed at the National Bureau of Standards to cover the range of pressures from 100 to 1 × 10−2 Torr. In these instruments the levels of the two liquid surfaces are measured by adjusting micrometers with conical points ground on the ends of the spindles. One hundred-millimeter and 50-mm range mercury manometers, with spindles approaching the liquid surface from above, are used for differential or absolute pressure measurements up to 100 Torr. Pressures less than 3.5 Torr are measured with oil manometers, in which the micrometer spindles approach the surfaces from below. Measurements made with the mercury manometer have an uncertainty of about 4 × 10−3 Torr plus 8 parts in 105 of the reading. Measurements made with the oil manometer have an uncertainty of about 4 × 10−4 Torr plus 1 part in 104 of the reading.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.