Abstract

The performance of a sensing module for pO 2, pCO 2, and pH was tested from a practical point of view and its design was modified for the pO 2 electrode and the pH combination electrode. Owing to the elongated lifetime of the on-chip liquid-junction reference electrode, pH measurement could be conducted repeatedly. The outputs from these electrodes taken in a 4.6 μl flow channel on the chip and in a 100 ml beaker were compared. Although a substantial difference was observed between the measured currents of the pO 2 electrode with a 0.2 mm×1.2 mm cathode, the difference was reduced markedly by using a 25 μm×25 μm cathode. A good coincidence was observed between the values of potential obtained with the pH electrode, while the data with the pCO 2 electrode suggested that it needed a substantial modification of design depending upon its applications. The sensing module was actually used to measure blood gas levels in whole human blood and the values were compared with those obtained by a commercial blood gas analyzer. The deviation of pH values was within 0.06 pH units and pO 2 values were within an acceptable level. However, substantial deviations of pCO 2 values were observed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.