Abstract

A planar, chip-based flow cell for UV-vis absorbance detection in HPLC is presented. The device features a microfabricated free-standing liquid core waveguide (LCW) capillary detection tube of long path length that is based on total internal reflection. We report on the linearity and calibration slope characteristics of lithographically produced LCWs with different interior/exterior geometries. 3D ray tracing was indispensable in modeling behavior in the more demanding geometries: multipath behavior may be intrinsic to these waveguides with consequent nonlinearity. Fortunately, nonlinearity in lithographically easy-to-produce waveguide geometries (such as with a flat, concave exterior and a round interior) is not as detrimental as might be initially expected. Experimental performance is predictably affected by the attainable surface quality of the LCW and efficient and reproducible coupling of the input light into the LCW.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.