Abstract

Simple, low-cost and portable microfluidic paper-based analytical devices (μPADs) for determination of creatinine in urine samples were developed. The methodology was based on Jaffé reaction between the creatinine and picric acid in alkaline conditions, generating a colorimetric creatinine-alkaline picrate complex. The product exhibits an orange color that is clearly visible on the μPADs. The color intensity of the complex, which is indicative of the concentration of creatinine, is then quantitatively determined using ImageJ software. Various experimental parameters were optimized to achieve the best performance of the μPADs. Under the optimum conditions, a wide linear range was obtained in the range of 0.2 - 1 mM with a limit of detection and limit of quantitation of 0.08 and 0.26 mM, respectively. The accuracy of the proposed method was in good agreement with the standard Jaffé method. Finally, the developed devices were successfully applied for the determination of creatinine in urine samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.