Abstract
We designed a microfluidic chip for use with an evanescent field optical sensor to conduct simultaneous measurements for ABO and Rh(D) forward blood typing. The microfluidic chip was designed to include multiplexed mixers for blood-reagent mixing and chambers for optical measurements. To achieve simple and rapid mixing in the microfluidic channels, holes containing freeze-dried reagents were employed as mixers. In addition, capillary stop valve-based multiplexing of the measurement chambers was used to effectively fill the chambers with samples without empty spaces or leaks. The five-channel microfluidic chip was designed and tested, including the mixing holes and measurement chambers. A multi-channel waveguide-mode sensor was used for optical measurement. Forward blood typing tests using human whole blood samples and anti-A, anti-B, anti-D, and Rh control reagents were conducted on the microfluidic chip. Simultaneous measurements were successfully performed using 40μL of diluted blood samples, and ABO and Rh(D) blood types could be determined within 5min.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.