Abstract

Optimization based on central composite design (CCD) for enantioseparation of anisodamine (AN), atenolol (AT), and metoprolol (ME) in human urine was developed using a microfluidic chip-CE device. Coupling the flexible and wide working range of microfluidic chip-CE device to CCD for chiral separation of AN, AT, and ME in human urine, a total of 15 experiments is needed for the optimization procedure as compared to 75 experiments using the normal one variable at a time optimization. The optimum conditions obtained are found to be more robust as shown by the curvature effects of the interaction factors. The developed microfluidic chip-CE-ECL system with adjustable dilution ratios has been validated by satisfactory recoveries (89.5-99% for six enanotiomers) in urine sample analysis. The working range (0.3-600 μM), repeatability (3.1-4.9% RSD for peak height and 4.0-5.2% RSD for peak area), and detection limit (0.3-0.6 μM) of the method developed are found to meet the requirements for bedside monitoring of AN, AT, and ME in patients under critical conditions. In summary, the hyphenation of CCD with the microfluidic chip-CE device is shown to offer a rapid means for optimizing the working conditions on simultaneous separation of three racemic drugs using the microfluidic chip-CE device developed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.