Abstract

Recent advances in medical radiation technologies have enabled small field radiotherapy with a beam size as small as a few millimeters for precise dose delivery to patients. However, conventional radiation dosimeters for small field radiotherapy suffer from inaccurate dose measurements due to their large absorber size. Herein, we present a nanofabricated chip calorimeter as a new platform for a miniaturized dosimeter for measurements of the absorbed dose. The calorimeter consists of a microscale absorber body built with a SU-8 photoresist and a microfluidic water container. The measurement system was constructed such that the calorimeter operates in the heat compensation mode, where the temperature change of the absorber was proportional to the absorbed energy. The operability of the heat compensation calorimeter was examined with electric heating and a visible light source. The feasibility of application of the microfluidic chip calorimeter for medical radiation sensor was successfully demonstrated by measuring the absorbed dose from the small radiation field of the Gamma Knife.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.