Abstract
This letter demonstrates a microfluidic platform for enumerating CD4+ T-lymphocytes from whole blood using chemiluminescence as a detection method. We microfabricated traps in a chamber and coated them with anti-CD4 antibody to isolate CD4+ T-cells. Based on cell surface-bound CD3 antibodies conjugated with horseradish peroxidase, incubation with chemiluminescent substrate produced a current in the photodetector that was proportional to the number of captured CD4+ T-cells. Analyzing 3 microL of whole blood, the platform exhibited high cell-capture efficiency and produced cell counts with high correlation to results obtained from flow cytometry. Compared to other lab-on-a-chip methods for CD4 counting, this method uses an instrument that requires no external light source and no image processing to produce a digitally displayed result only seconds after running the test.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.